PiKo-1
Specifications |
Range: |
PEF: 15
- 999 L/min (1 L/min resolution) FEV1: 0.15
- 9.99 Liter (0.01 Liter resolution) |
Accuracy: |
PEF: ±5%
or 20 L/min (whichever is greater) FEV1: ±3% or
0.1 Liter (whichever is greater) |
Sensor: |
Pressure/flow
sensor technology, (patented) |
Memory: |
96 Tests
contain: PEF, FEV1, Color Zone, Quality
Factor, Time/Date Stamp (15 min resolution) |
Memory
Type: |
Non-volatile
Color Zone: 3 Color Zones (Green, Yellow, Red) |
Reference
Values: |
User settable
for PEF or FEV1
Quality Factor: Warning & indicator for cough or abnormal blow |
Sounds: |
4 patterns
for different indications and warnings |
Communication: |
Bi-directional
IR port (RS232 format) |
External
Settings: |
Possible
settings using the optional software:
- Each Color Zone limits (in 10% increments)
- Select Color Zones to relate to PEF or FEV1 |
Battery
Life: |
2 year (based
on average of 6 tests per day), or
21,000 continuous blows |
Battery
Type: |
2 type 357
silver oxide button cells (or equivalent) |
Dimensions: |
75 x 35
x 20 mm |
Weight:
|
35 g |
Material
Type: |
Latex free |
Back
Pressure: |
1.5 cmH20/L/sec
at 14 L/sec or lower |
Operating
Temp: |
5 to 40oC |
Storage
Temp: |
(-20) to
70oC |
Humidity: |
20 - 95%
relative humidity |
Barometric:
|
550 to 780
mm Hg |
Standard: |
ATS 1994
(monitoring devices); waveforms 24 and 26 |
Compliance: |
AS/NZS-4237:
1994
EN60601, EN60601-1-1, EN60601-1-2 |
Regulatory:
|
FDA - 510(k)
for OTC
CE (0086) - Class I, with measurement function |
Warranty: |
6 months
(batteries not included) |